![]() ![]() To perform Open Compensation, disconnect the DUT terminals from the impedance analysis topology, leaving the DUT an open circuit. Performing Compensation: First, measure the reference resistors actual value with a precise multimeter and enter that value into the Compensation window text field under “Resistor”. ![]() If the resistor values or frequency range changes, this compensation must be performed again. When compensation is performed, values are stored for each reference resistor value over the specified frequency range. This is useful when the reference resistor value is not exactly it's marked value. These reference resistor values can also be changed in this window and reflected in the Control Bar's Resistor drop-down menu. This window provides some parameters found on the Control Bar as well as buttons for open/short compensation for multiple reference resistor options. Image B shows the Compensation configuration window shown when “Compensation” is clicked. Image A shows the Compensation tools icon selected, which allows the user to enable and perform open and short circuit compensation. ![]() This becomes more useful when working with higher frequencies, using higher reference resistor values, and detecting smaller capacitances and inductances. USB Scopes, Analyzers and Signal GeneratorsĬompensation helps calibrate the Impedance Analyzer. ![]()